Wideband acoustic immittance: normative study and test-retest reliability of tympanometric measurements in adults
Issue Date
Embargo End Date
Authors
Advisor
Citation
Sun, Xiao-Ming. 2016. Wideband acoustic immittance: normative study and test-retest reliability of tympanometric measurements in adults. Journal of Speech, Language, and Hearing Research, vol. 59:no.4:pp 819-834
Abstract
Purpose: The purpose of this study was to present normative data of tympanometric measurements of wideband acoustic immittance and to characterize wideband tympanograms.
Method: Data were collected in 84 young adults with strictly defined normal hearing and middle ear status. Energy absorbance (EA) was measured using clicks for 1/12-octave frequencies (0.236 to 8 kHz), with the ear canal air pressure systematically varied (+200 to -300 daPa). In 40 ears, 7 consecutive trials and a trial of clinical 226-Hz acoustic admittance (Y-a) tympanometry followed. A cavity test was also conducted.
Results: From the wideband EA tympanogram, several EA spectrums and EA tympanograms were derived. Descriptive statistics were performed, and population parameters were estimated. The immediate test-retest reliability was excellent. Effects of ear canal air pressure on EA were examined comprehensively. Differences in EA between tympanometric and ambient-pressure measurements were significant. Single-frequency EA tympanograms exemplified for half-octave frequencies were contrasted. The bandpass EA tympanogram, 0.236- and 1-kHz EA and Y-a tympanograms, and 226-Hz Y-a tympanogram were compared in 9 variables.
Conclusions: This study established a database of wideband tympanograms in healthy adults. The data analyses will promote our understanding of the middle ear transfer function. These data will serve as a reference for further studies in clinical populations.